BackBack

EVOXI - Rev. 3

EvominiX I

4÷20mA programmable temperature transmitter with mineral insulated probe

PC-configurable temperature transmitter with mineral insulated probe (MgO). It comes in a overmolded IP67 casing with M12 output connection. This device can be a reliable alternative to common assemblies with connection head.

  • Product Datasheet
  • Download
  • Videos

Technical Specifications

Electrical technical data
Electronic board operating temperature
-40 ÷85°C
Operating humidity
0 ÷100%
Power supply
5.5 ÷32 Vdc (polarity protected)
Input/Output insulation
None
Sensor input signal filter (*)
(*) time to reach 90% of signal
Configurable from 0.4 to 9.4 seconds
Signal
4÷20 mA
20÷4 mA
Permitted load
840Ω @ 24 Vdc [Rload= (Vsupply. - 5.5) / 0.022]
Sensor break or short-circuit monitoring
According to NAMUR NE43, selectable between:
Upper scale (> 21.0 mA)
Lower scale (< 3,6 mA)
Accuracy (*)
(*) @25°C
Maximum value between ±0.1K and ±0.1% of span
Temperature influence (*)
(*) deviation from 20°C
Maximum value between ±0,3°C/25°C and ±0,3% of span/25°C
Long-term stability
Maximum 0.1% of span per year
Linear error
Negligible
EMC
In accordance to EN 61326-1:2013 (CE)
In accordance to BS EN 61326-1:2013 (UKCA)
Range configurations
It is possible set the device by EVOMINI SET configuration kit (it is needed a PC with OS Windows).
Minimum span
20°C
Sensor error compensation
(±5°C) over two points
Measurement range
-50 ÷500°C
International protection marking (*)
(*) According to IEC 60529
IP67
Connection body material
THERMOPLASTIC
Sheet material
AISI 316L
Stem length L
150 mm
Other lenghts availabloe on request
Sheath diameter d
Ø 6 mm
Ø 3 mm
M.I.C. min. bending radius
3 times the outer diameter (except the sensing tip which length is ~30 mm)
Type of connector
male 4-pin connector with M12x1 metal screw lock (in accordance with IEC 61076-2-101 STANDARDS)
Pt100 sensor accuracy
Class A up to 300°C according to IEC 751
Response time
<3.5s for diameter 3mm, <13s for diameter 6mm (according to IEC 751)
Factory default
Range (0÷150)°C
(4÷20)mA output
Sensor break >21mA
Sensor short-circuit <3.6mA
Option
On request adjustment on 1 or 2 points
Extension cable in TPE with overmolded M12 connector (IP67). Available cable lengths: (3, 8 or 12)m (others on request)
Weight
14g (L=100mm e Ø=3mm)
28g (L=100mm e Ø=6mm)