You are now in our product catalog
Italiano | Deutsch | Français

EOSI - Rev. 2

EvominiSerI-RTD
Programmable temperature transmitter with mineral insulated probe and Modbus RTU digital output (RS485)
Temperature transmitter with overmolded IP67 casing with M12 output connection. By using a single cable, it is possible to build a sensors network which can be directly connected to a PLC or a PC equipped with a supervisory software (SCADA). These devices are equipped with M12 connectors allowing for an easy and quick installation with degree of protection IP67.
Need Help?
Sales Team
We will get back to you within 2 business days
Technical specifications
Electrical technical data
Power supply
9 ÷32 Vdc (polarity protected)
Electronic board input
RTD Pt100 / Pt1000 (α= 0,00385 / α= 0,003916) 2, 3 o 4 wire connection
Sensor exciting current
~100 uA
Sensor wire maximum resistance
2 wire connection: 40 ohm
3 or 4 wire connection: 20 ohm/wire
Accuracy (*)
(*) @25°C
Converter: ≤ ±0.2°C of the range
Sensor element: Pt100 class A up to 300 °C according to IEC751
Temperature influence (*)
(*) deviation from 20°C
< ±0.25°C/25°C of the range
Ambient temperature
-40 ÷80°C
Resolution
0,1 °C
Linear error
Negligible
Sensor error compensation
(±5°C) over two points
Current consumption
<4mA with RS485 (<10mA at power on for approx. 8mS)
Communication protocol
Modbus RTU (max. baud rate 38.400)
Serial interface
RS-485 (not insulated)
Input/Output insulation
None
Maximum connection distance
1000 meters, this distance depends on the power supply voltage and the type of cable used to connect the various devices
Maximum devices on network
maximum 32 (over is required to add a serial repeater)
Maximum 256 nodes (it is recommended to use isolators / repeaters along the serial line)
Indicator LED
Blue LED, power supply and device operations indicator
Red LED (ERR), sensor error indicator
White LEDs (TX and RX), serial transmission and reception indicators
International protection marking (*)
(*) According to IEC 60529
IP65/67
EMC
In accordance to EN 61326-1:2013 (CE)
In accordance to BS EN 61326-1:2013 (UKCA)
Configuration
By using the EVOPLATFORMSET configuration kit (a PC with Windows OS is required).
Main device design parameters
Maximum and minimum temperature peak measured
Temperature offset for measurement correction
Serial communication and power-on watch-dog
Temperature tenths °C
Measurement range
-50 ÷350°C
Material body
Thermoplastic
Type of connector
1 input connector M12 x 4 male according to IEC 61076-2-101 STANDARDS (power supply / serial interface)
Sensor construction type
with mineral compact insulation (M.I.C.)
Stem length L
150 mm
250 mm
350 mm
Other lengths on request
Dimensional notes
Lengths other than those listed can be produced for minimum quantities to be agreed (after our feasibility study)
Sheet material
AISI 316L
Sheath diameter d
Ø 6 mm
Ø 3 mm
M.I.C. min. bending radius
3 times the outer diameter (except the sensing tip which length is ~30 mm)
Insulation resistance
100 M Ω@ 100 Vdc.
Response time (*)
(*) test in water in accordance with IEC 751. Time taken to reach 63.2% of temperature step
less than 3.5 seconds for Ø 3 mm and less than 13 seconds for diameter Ø 6 mm
Weight
14g (L=100mm e Ø3mm); 28g (L=100mm e Ø6mm)
28g (L=100mm e Ø=6mm)
Option
"T" distributor Female / Male / Female M12 x 5 poles
Extension cables with M12 overmoulded female and male connectors
Wall mounting bracket
EVOPLATFORMSET configuration kit
Media